The gas detector Pac 8000 NO reliably and accurately detects the concentration of ammonia and warns of dangerous gas concentrations.

  • Single gas measuring device for personal measurement
  • XXS sensor technology with high response speed and membrane filter
  • Standard measuring range: 0 - 200 ppm, A1/A2 alarm thresholds: 20/40 ppm
  • 360° alarm signal, acoustic (90 dB) and optical
  • The gas detector features a robust, non-slip housing for use under extreme conditions
  • Large and user-friendly display with bright backlighting
  • Storage of concentrations and events with date and time

 

Required function tests before each use should be performed with the Dräger test and calibration stations as well as the associated test gases. Basis for the function tests are the information sheets T021 / T023 (DGUV Information 213-056 or DGUV Information 213-057). Further information can be found here.

If you wish for a custom alarm threshold configuration, please specify the alarm thresholds in the comment field at the end of the order process.

Further information about the product can be found on our website. Please follow the link.


Technical Data - Dräger Sensor XXS NO

Detection Limit 0,3 ppm
Resolution 0,1 ppm
Measuring Range 0 - 200 ppm NO (Nitric Oxide)
Response Time ≤ 10 seconds (t90)
Precision Sensitivity ≤ ± 3 % of the measured value
Long-term Drift (at 20 °C)
  • Zero Point: ≤ ± 0,3 ppm/year
  • Sensitivity: ≤ ± 2 % of the measured value/month
Warm-up Time ≤ 20 hours
Environmental Conditions
  • Temperature: (-40 to 50) °C (-40 to 122) °F
  • Humidity: (10 to 90) % RH
  • Pressure: (700 to 1300) hPa
Temperature Influence
  • Zero Point: ≤ ± 0,02 ppm/K
  • Sensitivity: ≤ ± 0,3 % of the measured value/K
Humidity Influence
  • Zero Point: no influence
  • Sensitivity: ≤ ± 0.05 % of the measured value/% r. F.
Test Gas approx. 3 to 1 ppm NH3

Relevant Cross Sensitivities - Dräger Sensor XXS NO

The values given in the table are guideline values and apply to new sensors. The values specified can vary by ±30%. The sensor may also be sensitive to other gases (data available upon request from Dräger). Gas mixtures can be displayed as a sum. Gases with negative sensitivity can negate a positive NH3 display. It should be checked whether gas mixtures are present.

Gas/Vapor Concentration Display in ppm NO
Acetone (CH3COCH3)1000ppmno influence
Ammonia (NH3) 500 ppmno influence
Benzene (C6H6) 0,6 Vol.-% no influence
Chlorine (Cl2) 5 ppmno influence
Carbon monoxide (CO) 2000 ppm no influence
Methane (CH4) 2 Vol.-% no influence
Hydrogen sulphide (H2S) 5 ppm 1
Nitrogen dioxide (NO2) 20 ppm no influence
Hydrogen (H2) 1,5 Vol.-% no influence


(-) negative deviation sign

Rental Advanced: Yes
Rental item or Consumable?: Rental items