The Pac 8000 CO2 gas warning device reliably and accurately detects the concentration of carbon dioxide and warns of dangerous gas concentrations.

  • Single gas measuring device for personal measurement
  • XXS sensor technology with high response speed and membrane filter
  • Standard measuring range: 0 - 5 vol.-%, A1/A2 alarm thresholds: 0.5/1 vol-%
  • 360° alarm signal, acoustic (90 dB) and optical
  • The CO2 gas warning device has a robust, grippy housing for use in extreme conditions
  • Large and user-friendly display with bright backlighting
  • Storage of concentrations and events with date and time

 

Carry out the prescribed functional tests before each use with the Dräger test and calibration stations and the associated test gases. The basis for the functional tests are the information sheets T021 / T023 (DGUV Information 213-056 or DGUV Information 213-057). Further information can be found here.

If you wish to have an individual alarm threshold configuration, please specify the alarm thresholds at the end of the order process in the comment field.

Further information about the product can be found on our website. Please follow the link.


Technical Data - Dräger Sensor XXS CO2

Detection Limit 0.3 vol.-%
Resolution 0.1 vol.-%
Measuring Range 0 to 5 vol.-% CO2 (carbon dioxide)
Response Time ≤ 30 seconds (t50)
Precision Sensitivity ≤ ± 20 % of the measured value
Long-term Drift (at 20 °C)
  • Zero point: ≤ ± 0.2 vol.-%/year
  • Sensitivity: ≤ ± 15 % of the measured value/month
Warm-up Time ≤ 12 hours
Ambient Conditions
  • Temperature: (-20 to 40) °C (-4 to 104) °F
  • Humidity: (10 to 90) % r.h.
  • Pressure: (700 to 1300) hPa
Temperature Influence
  • Zero point: ≤ ± 0.01 vol.-%/K
  • Sensitivity: ≤ ± 2 % of the measured value/K
Humidity Influence
  • Zero point: no influence
  • Sensitivity: ≤ ± 0.1 % of the measured value/% r.h.
Test Gas 1 to 4 vol.-% CO2

Relevant Cross-Sensitivities - Dräger Sensor XXS CO2

The values given in the table are guidelines and apply to new sensors. The specified values may vary by ±30%. The sensor may also be sensitive to other gases (data available on request from Dräger). Gas mixtures can be displayed as a total. Gases with negative sensitivity can offset a positive CO2 reading. It should be checked whether gas mixtures are present.

Gas/Vapor Concentration Display in ppm CO2
Ammonia (NH3) 50 ppm no influence
Chlorine (Cl2) 10 ppm no influence
Hydrochloric Acid (HCl) 20 ppm no influence
Hydrogen Cyanide (HCN) 60 ppm no influence
Ethanol (C2H5OH) 250 ppm no influence
Ethylene (C2H2) 100 ppm no influence
Isobutene ((CH3)2CCH2) 100 ppm no influence
Carbon Monoxide (CO) 1000 ppm no influence
Methane (CH4) 0.9 vol.-% no influence
Ozone (O3) 1.5 ppm no influence
Phosphine (PH3) 5 ppm no influence
Sulfur Dioxide (SO2) 20 ppm no influence
Hydrogen Sulfide (H2S) 20 ppm no influence
Nitrogen Dioxide (NO2) 20 ppm no influence
Nitric Oxide (NO) 20 ppm no influence
Hydrogen (H2) 1.6 vol.-% no influence

(-) negative sign of deviation

Rental Advanced: Yes
Rental item or Consumable?: Rental items
Sensors: CO2
Skip product gallery

Accessory Items

A metal gas cylinder with a Dräger label displaying calibration gas and additional information
Calibration Gas CO2, 50%vol in N2
5239084

$318.07*
Skip product gallery

Similar Items

Calibration station with display and holder for a Dräger X-dock 5300 device and inserted H2S gas monitor, black and gray color scheme
X-dock Set 5300 for the Pac Series
SRM09205_US

From $32.47* per Day
Black test station for gas detectors by Dräger with holders and a label showing instructions
Bump Test Station Pac 1000-8000 US
SRM06768

$0.00* per Day
Black communication module with long USB cable on white background
Communication Module for Pac (USB-DIRA)
KIT46_US

From $3.26* per Day