The gas warning device Pac 8000 Cl2 reliably and accurately detects the concentration of chlorine gas and warns at dangerous gas concentrations.

  • Single gas measuring device for personal measurement
  • XXS sensor technology with high reaction speed and membrane filter
  • Standard measurement range: 0 - 20 ppm, A1/A2 alarm thresholds: 0.5/1 ppm
  • 360° alarm signal, acoustic (90 dB) and visual
  • The gas detector features a robust, handy housing for use under extreme conditions
  • Large and user-friendly display with bright backlighting
  • Storage of concentrations and events with date and time

 

Conduct the prescribed function tests before each use with the Dräger test and calibration stations and the associated test gases. Basis for the function tests are the information sheets T021 / T023 (DGUV Information 213-056 or DGUV Information 213-057). Further information can be found here.

If you wish for an individual alarm threshold configuration, please specify the alarm thresholds at the end of the ordering process in the comment field.

Further information about the product can be found on our website. Please follow the link.


Technical Data - Dräger Sensor XXS Cl2

Detection limit 0.05 ppm
Resolution 0.05 ppm
Measurement range/Relative sensitivity
  • 0 to 20 ppm Cl2 (Chlorine) → 1.00
  • 0 to 20 ppm F2 (Fluorine) → 1.00
  • 0 to 20 ppm Br2 (Bromine) → 1.00
  • 0 to 20 ppm ClO2 (Chlorine dioxide) → 0.60
Response time ≤ 30 seconds (t90)
Precision sensitivity ≤ ± 2 % of the measured value
Long-term drift (at 20 °C)
  • Zero point: ≤ ± 0.2 ppm/year
  • Sensitivity: ≤ ± 2 % of the measured value/month
Stabilization time ≤ 30 minutes
Environmental conditions
  • Temperature: (-40 to 50) °C (-40 to 122) °F
  • Humidity: (10 to 90) % rH
  • Pressure: (700 to 1300) hPa
Temperature influence
  • Zero point: ≤ ± 0.05 ppm
  • Sensitivity: ≤ ± 5 % of the measured value
Humidity influence
  • Zero point: no influence
  • Sensitivity: ≤ ± 0.4 % of the measured value/% rH
Test gas approx. 1 to 18 ppm Cl2

Relevant Cross Sensitivities - Dräger Sensor XXS Cl2

The values stated in the table are indicative and apply to new sensors. The indicated values may vary by ±30%. The sensor can also be sensitive to other gases (data upon request from Dräger). Gas mixtures can be displayed as a sum. Gases with negative sensitivity can offset a positive display of Cl2. It should be checked whether gas mixtures are present.

Gas/Vapor Concentration Display in ppm Cl2
Ammonia (NH3) 50 ppm no influence
Hydrogen chloride (HCl) 20 ppm ≤ 0.5
Hydrogen cyanide (HCN) 60 ppm no influence
Ethanol (C2H5OH) 250 ppm no influence
Acetylene (C2H2) 100 ppm no influence
Isobutylene ((CH3)2CCH2) 100 ppm no influence
Carbon dioxide (CO2) 10 vol.-% no influence
Carbon monoxide (CO) 1000 ppm no influence
Methane (CH4) 0.9 vol.-% no influence
Ozone (O3) 1 ppm no influence
Phosphine (PH3) 1 ppm no influence
Sulfur dioxide (SO2) 10 ppm ≤ 1 (–)
Hydrogen sulfide (H2S) 10 ppm ≤ 0.6 (–)
Nitrogen dioxide (NO2) 10 ppm no influence
Nitric oxide (NO) 20 ppm no influence
Hydrogen (H2) 1000 ppm no influence

(-) negative sign of deviation

Rental Advanced: Yes
Rental item or Consumable?: Rental items
Sensors: CI2

Accessory Items

Calibration Gas CO, 100 / O2, 18 / N2, 112
SRM10633