Pac 8000 Single Gas Monitor for Chlorine Gas (CL2)
The gas warning device Pac 8000 Cl2 reliably and accurately detects the concentration of chlorine gas and warns at dangerous gas concentrations.
- Single gas measuring device for personal measurement
- XXS sensor technology with high reaction speed and membrane filter
- Standard measurement range: 0 - 20 ppm, A1/A2 alarm thresholds: 0.5/1 ppm
- 360° alarm signal, acoustic (90 dB) and visual
- The gas detector features a robust, handy housing for use under extreme conditions
- Large and user-friendly display with bright backlighting
- Storage of concentrations and events with date and time
Conduct the prescribed function tests before each use with the Dräger test and calibration stations and the associated test gases. Basis for the function tests are the information sheets T021 / T023 (DGUV Information 213-056 or DGUV Information 213-057). Further information can be found here.
If you wish for an individual alarm threshold configuration, please specify the alarm thresholds at the end of the ordering process in the comment field.
Further information about the product can be found on our website. Please follow the link.
Technical Data - Dräger Sensor XXS Cl2
| Detection limit | 0.05 ppm |
|---|---|
| Resolution | 0.05 ppm |
| Measurement range/Relative sensitivity |
|
| Response time | ≤ 30 seconds (t90) |
| Precision sensitivity | ≤ ± 2 % of the measured value |
| Long-term drift (at 20 °C) |
|
| Stabilization time | ≤ 30 minutes |
| Environmental conditions |
|
| Temperature influence |
|
| Humidity influence |
|
| Test gas | approx. 1 to 18 ppm Cl2 |
Relevant Cross Sensitivities - Dräger Sensor XXS Cl2
The values stated in the table are indicative and apply to new sensors. The indicated values may vary by ±30%. The sensor can also be sensitive to other gases (data upon request from Dräger). Gas mixtures can be displayed as a sum. Gases with negative sensitivity can offset a positive display of Cl2. It should be checked whether gas mixtures are present.
| Gas/Vapor | Concentration | Display in ppm Cl2 |
|---|---|---|
| Ammonia (NH3) | 50 ppm | no influence |
| Hydrogen chloride (HCl) | 20 ppm | ≤ 0.5 |
| Hydrogen cyanide (HCN) | 60 ppm | no influence |
| Ethanol (C2H5OH) | 250 ppm | no influence |
| Acetylene (C2H2) | 100 ppm | no influence |
| Isobutylene ((CH3)2CCH2) | 100 ppm | no influence |
| Carbon dioxide (CO2) | 10 vol.-% | no influence |
| Carbon monoxide (CO) | 1000 ppm | no influence |
| Methane (CH4) | 0.9 vol.-% | no influence |
| Ozone (O3) | 1 ppm | no influence |
| Phosphine (PH3) | 1 ppm | no influence |
| Sulfur dioxide (SO2) | 10 ppm | ≤ 1 (–) |
| Hydrogen sulfide (H2S) | 10 ppm | ≤ 0.6 (–) |
| Nitrogen dioxide (NO2) | 10 ppm | no influence |
| Nitric oxide (NO) | 20 ppm | no influence |
| Hydrogen (H2) | 1000 ppm | no influence |
(-) negative sign of deviation
| Rental Advanced: | Yes |
|---|---|
| Rental item or Consumable?: | Rental items |
| Sensors: | CI2 |