The gas warning device Pac 8000 NO2 reliably and accurately detects the concentration of nitrogen dioxide and warns of dangerous gas concentrations.

  • Single gas measuring device for personal measurement
  • XXS sensor technology with high reaction speed and membrane filter
  • Standard measuring range: 0 - 50 ppm, A1/A2 alarm thresholds: 5/10 ppm
  • 360° alarm signal, acoustic (90 dB) and visual
  • The gas warning device has a robust, grippy housing for use under extreme conditions
  • Large and user-friendly display with bright backlight
  • Storage of concentrations and events with date and time

 

Conduct the prescribed functional tests before each use with the Dräger test and calibration stations and the associated test gases. The basis for the functional tests are the information sheets T021 / T023 (DGUV Information 213-056 or DGUV Information 213-057). Further information can be found here.

If you wish to configure individual alarm thresholds, please indicate the alarm thresholds in the comment field at the end of the order process.

For further information about the product, visit our website. Please follow the link.


Technical Data - Dräger Sensor XXS NO2

Detection Limit 0.04 ppm
Resolution 0.02 ppm
Measuring Range 0 to 50 ppm NO2 (nitrogen dioxide)
Response Time ≤ 15 seconds (t50)
Precision Sensitivity ≤ ± 3 % of measured value
Long-term Drift (at 20 °C, 50% RH)
  • Zero point: ≤ ± 0.04 ppm/year
  • Sensitivity: ≤ ± 2 % of measured value/month
Run-in Time ≤ 120 minutes
Environmental Conditions
  • Temperature: (-30 to 50) °C (-22 to 122) °F
  • Humidity: (15 to 80) % RH
  • Pressure: (700 to 1300) hPa
Temperature Influence
  • Zero point: no influence
  • Sensitivity: ≤ ± 0.5 % of measured value/K
Humidity Influence
  • Zero point: no influence
  • Sensitivity: ≤ ± 0.1 % of measured value/% RH
Test Gas approx. 0.5 to 45 ppm NO2

Relevant Cross Sensitivities - Dräger Sensor XXS NO2

The values given in the table are guidelines and apply to new sensors. The specified values may vary by ±30%. The sensor may also be sensitive to other gases (data on request from Dräger). Gas mixtures can be displayed as a sum. Gases with negative sensitivity can negate a positive display of NO2. It should be checked whether gas mixtures are present.

Gas/Vapor Concentration Display in ppm NO2
Ammonia (NH3) 30 ppm no influence
Arsine (AsH3) 0.5 ppm no influence
Chlorine (Cl2) 1 ppm ≤ 1.5
Chlorine dioxide (ClO2) 1 ppm ≤ 1.5
Sulfur dioxide (SO2) 1 ppm ≤ 0.12 (–)
Hydrogen sulfide (H2S) 1 ppm ≤ 0.03 (–)
Ozone (O3) 0.5 ppm ≤ 1
Carbon monoxide (CO) 2000 ppm no influence

(-) negative deviation sign