The Pac 8000 Phosgene gas detection device reliably and accurately detects the concentration of phosgene and alerts at dangerous gas concentrations.

  • Single gas measurement device for personal measurement
  • XXS sensor technology with high response speed and membrane filter
  • Standard measurement range: 0 - 10 ppm, A1/A2 alarm thresholds: 0.1/0.2 ppm
  • 360° alarm signal, acoustic (90 dB) and visual
  • The gas detector has a robust, easy-grip housing for use under extreme conditions
  • Large and user-friendly display with bright backlight
  • Storage of concentrations and events with date and time

 

Mandatory function tests before each use are conducted with the Dräger test and calibration stations as well as the corresponding test gases. Basis for the function tests are the information sheets T021 / T023 (DGUV Information 213-056 or DGUV Information 213-057). Further information can be found here.

If you wish for an individual alarm threshold configuration, please specify the alarm thresholds in the comment field at the end of the order process.

Further information about the product can be found on our website. Please follow the link.


Technical Data - Dräger Sensor XXS COCl2

Detection limit 0.01 ppm
Resolution 0.01 ppm
Measurement range 0 to 10 ppm COCl2 (phosgene)
Response time ≤ 20 seconds (t20)
Precision sensitivity ≤ ± 5 % of the measured value
Long-term drift (at 20 °C)
  • Zero point: ≤ ± 0.01 ppm/year
  • Sensitivity: ≤ ± 1 % of the measured value/month
Warm-up time ≤ 1 hour
Ambient conditions
  • Temperature: (-20 to 35) °C (-4 to 99) °F
  • Humidity: (10 to 90) % relative humidity
  • Pressure: (700 to 1300) hPa
Storage (+4 … +8°C)
Temperature influence
  • Zero point: no influence
  • Sensitivity: ≤ ± 0.2 % of the measured value/K
Humidity influence
  • Zero point: no influence
  • Sensitivity: ≤ ± 0.05 % of the measured value/% relative humidity
Test gas COCl2 test gas between 3.8 to 9 ppm (not in the Dräger program).

Relevant Cross Sensitivities - Dräger Sensor XXS COCl2

The values indicated in the table are guidelines and apply to new sensors. The indicated values may vary by ±30%. The sensor may also be sensitive to other gases (data on request from Dräger). Gas mixtures can be displayed as a sum. Gases with negative sensitivity can negate a positive display of COCl2. It should be checked whether gas mixtures are present.

Gas/Vapor Concentration Display in ppm COCl2
Ammonia (NH3) 20 ppm no influence
Chlorine (Cl2) 0.5 ppm ≤ 0.2
Hydrogen chloride (HCl) 0.5 ppm ≤ 0.7
Hydrogen cyanide (HCN) 5 ppm ≤ 6.5
Ethanol (C2H5OH) 260 ppm no influence
Ethylene (C2H2) 20 ppm no influence
Hydrogen fluoride (HF) 0.4 ppm ≤ 0.1 ppm
Isobutene ((CH3)2CCH2) 100 ppm no influence
Carbon dioxide (CO2) 1.5 vol.-% no influence
Carbon monoxide (CO) 1000 ppm no influence
Ozone (O3) 0.3 ppm ≤ 0.05 (–)
Phosphine (PH3) 0.5 ppm ≤ 0.1 ppm
Propanol (C3H7OH) 500 ppm no influence
Sulfur dioxide (SO2) 2 ppm no influence
Hydrogen sulfide (H2S) 1 ppm ≤ 1
Nitrogen dioxide (NO2) 1 ppm ≤ 0.1 (–)
Nitric oxide (NO) 30 ppm no influence
Hydrogen (H2) 8000 ppm no influence
Hydrogen peroxide (H2O2) 1 ppm no influence

(-) negative sign of deviation

Rental Advanced: Yes
Rental item or Consumable?: Rental items
Sensors: Phosgen