The gas warning device Pac 8000 HCN reliably and accurately detects the concentration of hydrogen cyanide and warns of dangerous gas concentrations.

  • Single-gas measuring device for personal measurement
  • XXS sensor technology with high reaction speed and membrane filter
  • Standard measuring range: 0 - 50 ppm, A1/A2 alarm thresholds: 1.9/3.8 ppm
  • 360° alarm signal, acoustic (90 dB) and visual
  • The gas detector has a robust, handy casing for use in extreme conditions
  • Large and user-friendly display with bright backlighting
  • Storage of concentrations and events with date and time

 

Conduct mandatory function tests before each use with the Dräger test and calibration stations and the associated test gases. The basis for the function tests are the information sheets T021 / T023 (DGUV Information 213-056 or DGUV Information 213-057). Further information can be found here.

If you wish for an individual alarm threshold configuration, please specify the alarm thresholds at the end of the ordering process in the comment field.

Further information about the product can be found on our website. Please follow the link.


Technical Data - Dräger Sensor XXS HCN

Detection limit 0.5 ppm
Resolution 0.1 ppm
Measuring range 0 to 50 ppm HCN (hydrogen cyanide)
Response time ≤ 10 seconds (t50)
Precision Sensitivity ≤ ± 5 % of the measured value
Long-term drift (at 20 °C)
  • Zero point: ≤ ± 2 ppm/year
  • Sensitivity: ≤ ± 5 % of the measured value/month
Run-in time ≤ 35 minutes
Environmental conditions
  • Temperature: (-20 to 50) °C (-4 to 122) °F
  • Humidity: (10 to 90) % r.H.
  • Pressure: (700 to 1300) hPa
Temperature influence
  • Zero point: ≤ ± 1 ppm
  • Sensitivity: ≤ ± 5 % of the measured value
Humidity influence
  • Zero point: no influence
  • Sensitivity: ≤ ± 0.1 % of the measured value/% r.H.
Test gas approx. 1 to 45 ppm HCN

Relevant Cross-Sensitivities - Dräger Sensor XXS HCN

The values given in the table are guidelines and apply to new sensors. The specified values can fluctuate by ±30%. The sensor can also be sensitive to other gases (data on request from Dräger). Gas mixtures can be displayed as a sum. Gases with negative sensitivity can cancel a positive display of HCN. It should be checked whether gas mixtures are present.

Gas/Vapor Concentration Display in ppm HCN
Ammonia (NH3) 50 ppm no influence
Chlorine (Cl2) 10 ppm ≤ 20 (–)
Hydrochloric acid (HCl) 20 ppm ≤ 1
Ethanol (C2H5OH) 250 ppm no influence
Ethylene (C2H2) 100 ppm ≤ 10
Isobutylene ((CH3)2CCH2) 100 ppm ≤ 1.5
Carbon dioxide (CO2) 10 Vol.-% no influence
Carbon monoxide (CO) 200 ppm no influence
Methane (CH4) 1 Vol.-% no influence
Ozone (O3) 0.5 ppm no influence
Phosgene (COCl2) 5 ppm no influence
Phosphine (PH3) 1 ppm ≤ 8
Sulfur dioxide (SO2) 20 ppm ≤ 10
Hydrogen sulfide (H2S) 20 ppm ≤ 50
Nitrogen dioxide (NO2) 10 ppm ≤ 20 (–)
Nitric oxide (NO) 20 ppm no influence
Hydrogen (H2) 1.5 Vol.-% ≤ 10

(-) negative sign of deviation

Rental Advanced: Yes
Rental item or Consumable?: Rental items
Sensors: HCN

Accessory Items

Calibration Gas C4H8, 100ppm in Air
SRM08009

$583.58*