The gas detection device Pac 8500 CO-H2 CP selectively measures carbon monoxide up to a hydrogen concentration of 1000 ppm (2.5% LEL).

Carbon monoxide and hydrogen can simultaneously occur in work areas such as the steel industry, refineries, and wastewater treatment plants. Hydrogen affects the CO signal in commercial sensors, leading to false alarms. The DrägerSensor® XXS CO H2-CP operates with two measurement electrodes, where one electrode measures CO and H2, and the other only H2. By calculating the difference between the two signals, only the CO value is displayed. Thus, a hydrogen concentration of 1000 ppm (2.5% LEL) results in a maximum display of 15 ppm CO without triggering the CO alarm.

  • Single-gas measuring device for personal measurement
  • Sensor: hydrogen-compensated carbon monoxide sensor
  • Standard measuring range: 0 to 2000 ppm
  • A1/A2 alarm thresholds: 20/40 ppm
  • Highly visible 360° alarm, acoustic and visual
  • The gas detector has a powerful XXS sensor technology with high response speed
  • Storage of concentrations and results with date and time

Perform mandatory function tests before each use with the Dräger test and calibration stations and the corresponding test gases. Basis for the function tests are the information sheets T021 / T023 (DGUV Information 213-056 or DGUV Information 213-057). For further information, please click here.

If you wish for an individual alarm threshold configuration, please specify the alarm thresholds at the end of the ordering process in the comment field.

Further information on the product can be found on our website. Please follow the link.


Technical Data - Dräger Sensor XXS CO H₂-CP

Detection Limit 6 ppm
Resolution 2 ppm
Measuring Range 0 to 2000 ppm CO (Carbon Monoxide)
Response Time ≤ 25 seconds (t₉₀)
Precision Sensitivity ≤ ± 2% of the measured value
Long-term Drift (at 20 °C)
  • Zero Point: ≤ ± 2% ppm/year
  • Sensitivity: ≤ ± 1% of the measured value/month
Settling Time ≤ 12 hours
Ambient Conditions
  • Temperature: (-20 to 50) °C (-4 to 122) °F
  • Humidity: (10 to 90) % r.h.
  • Pressure: (700 to 1300) hPa
Temperature Influence
  • Zero Point: ≤ ± 5 ppm
  • Sensitivity: ≤ ± 0.3% of the measured value/K
Humidity Influence
  • Zero Point: No influence
  • Sensitivity: ≤ ± 0.02% of the measured value/% r.h.
Test Gas 20 to 1800 ppm CO and 1000 ppm H2

Relevant Cross Sensitivities - Dräger Sensor XXS CO H₂-CP

The values given in the table are guide values and apply to new sensors. The values can vary by ±30%. The sensor may also be sensitive to other gases (data available on request from Dräger). Gas mixtures can be displayed as a sum. Gases with negative sensitivity can compensate for a positive CO display. It should be checked if gas mixtures are present.

Gas/Vapor Concentration Display in ppm CO with Selective Filter
Ammonia (NH₃) 100 ppm No influence
Chlorine (Cl₂) 20 ppm No influence
Hydrochloric Acid (HCl) 40 ppm No influence
Hydrogen Cyanide (HCN) 50 ppm No influence
Ethanol (C₂H₅OH) 250 ppm No influence
Ethylene (C₂H₄) 100 ppm ≤ 300
Ethylene (C₂H₂) 100 ppm ≤ 200
Isobutylene (CH₃)₂CCH₂ 100 ppm No influence
Carbon Dioxide (CO₂) 30 vol.% No influence
Methane (CH₄) 5 vol.% No influence
Propane (C₃H₈) 1 vol.% No influence
Sulfur Dioxide (SO₂) 25 ppm No influence
Hydrogen Sulfide (H₂S) 30 ppm No influence
Nitrogen Dioxide (NO₂) 20 ppm No influence
Nitric Oxide (NO) 30 ppm ≤ 5
Hydrogen (H₂) 0.1 vol.% ≤ ± 15¹)

(-) Negative sign of deviation

Rental Advanced: Yes
Rental item or Consumable?: Rental items
Sensors: CO