The gas warning device Pac 8500 O2/CO measures oxygen and carbon monoxide simultaneously and warns of dangerous gas concentrations.

  • 2- in single-gas device for personal measurement
  • Sensor combination: Oxygen and Carbon Monoxide
  • Standard measuring range: 0 - 25 Vol.-% O2, 0 - 2,000 ppm CO
  • A1/A2 alarm thresholds: 19/23 Vol.% O2, 20/40 ppm CO
  • Highly visible 360° alarm, acoustic and optical
  • The gas detector features powerful XXS sensor technology with high response speed
  • Low downtime due to simultaneous measurement of two gases
  • Storage of concentrations and results with date and time

 

Perform the prescribed functional tests before each use with the Dräger test and calibration stations and the associated test gases. Basis for the functional tests are the information sheets T021 / T023 (DGUV Information 213-056 or DGUV Information 213-057). Further information can be found here.

If you desire a custom alarm threshold configuration, please specify the alarm thresholds in the comment field at the end of the order process.

Further information about the product can be found on our website. Please follow the link.


Technical Data - Dräger Sensor XXS O2/CO LC

Detection limit 0.1 Vol.-% O2 / 1 ppm CO
Resolution 0.1 Vol.-% O2 / 1 ppm CO
Measuring range 0 to 25 Vol.-% O2 / 0 to 2000 ppm CO
Response time ≤ 15 seconds (t90)
Accuracy sensitivity O2: ≤ ± 1 % of the measured value, CO: ≤ ± 2 % of the measured value
Long-term drift (at 20 °C)
  • O2: Zero point ≤ ± 0.5 Vol.-%/year, CO: Zero point ≤ ± 2 ppm/year
  • Sensitivity: O2 ≤ ± 1 % of the measured value/year, CO ≤ ± 3 % of the measured value/year
Warm-up time O2: ≤ 15 minutes, CO: ≤ 15 minutes
Ambient conditions
  • Temperature: (-40 to 50) °C (-40 to 122) °F
  • Humidity: (10 to 90) % r. H.
  • Pressure: (700 to 1300) hPa
Temperature influence
  • O2: Zero point ≤ ± 0.2 Vol.-%
  • Sensitivity: O2 ≤ ± 2 % of the measured value, CO ≤ ± 0.3 % of the measured value/K
Humidity influence
  • O2: Zero point no influence
  • Sensitivity: O2 ≤ ± 0.1 % of the measured value/% r.H., CO ≤ ± 0.02 % of the measured value/% r.H.
Test gas approx. 12 to 20 Vol.-% O2, 20 to 1800 ppm CO

Relevant Cross-sensitivities - Dräger Sensor XXS O2/CO LC

The values given in the table are guidelines and apply to new sensors. The given values can vary by ±30%. The sensor can also be sensitive to other gases (data on request from Dräger). Gas mixtures can be displayed as a sum. Gases with negative sensitivity can negate a positive display of O2 or CO. It should be checked if gas mixtures are present.

Gas/Vapor Concentration Display in ppm CO/O2
Ammonia (NH3) 100 ppm no influence
Chlorine (Cl2) 20 ppm no influence
Ethylene (C2H4) 2 Vol.-% O2 ≤ 2(–), CO ≤ 250
Helium (He) 20 Vol.-% O2 ≤ 3*
Carbon Dioxide (CO2) 10 Vol.-% O2 no influence, CO ≤ 2
Hydrogen (H2) 0.5 Vol.-% O2 ≤ 0.8(–), CO ≤ 1000

(-) negative sign of the deviation

Rental Advanced: Yes
Rental item or Consumable?: Rental items
Sensors: CO, O2