Skip to main content Skip to search Skip to main navigation

The DrägerSensor XXS CO H2-CP measures carbon monoxide in areas where hydrogen is also present

  • No typical influence of hydrogen on CO measurement, due to two measuring electrodes built-in (1. CO and H2; 2. H2) that calculate the true CO content
    • For example, a hydrogen concentration of 1000 ppm (2.5% LEL) leads to a maximum display of 15 ppm CO without triggering the CO alarm
  • High selectivity thanks to a built-in internal filter that filters out accompanying gases such as alcohol or acidic gases
  • Usable in Dräger Single Gas Detector Pac 8500 and Dräger Multi Gas Detector X-am 5000/5600/8000 devices
  • Expected sensor lifespan of > 3000 years
  • Measurement range
    • 0 to 2000 ppm CO (carbon monoxide)
  • Alarm thresholds
    • A1 = 30 ppm; A2 = 60 ppm

Suitable Market Segments:
Steel industry, refineries, wastewater treatment plants

 

 

Rental Advanced: Yes
Rental item or Consumable?: Rental items
Sensors: CO

Accessory Items

Calibration Gas CO2, 1.5%vol in N2
SRM08024