DraegerSensor IR CO2 UK
6812190_UK
Rental fee
per
Day
£0.00*
The DrägerSensor IR CO2 precisely measures the concentration of carbon dioxide and warns of hazardous gas concentrations.
- Compatible with the Dräger X-am 5600 and Dräger X-am 8000
- Works on the principle of infrared absorption and offers high stability in daily use
- Fast response time in both diffusion and pump mode
- Measurement of CO2 in ambient air; measuring range 0 to 5 vol.% CO2
Technical data – DrägerSensor IR CO2
| Resolution | 0.01 vol.% CO2 or 50 ppm CO2, depending on the measuring range |
|---|---|
| Measuring range | 0 to 5 vol.% CO2 |
| Accuracy (sensitivity) |
|
| Long-term drift at 20 °C (68 °F) |
|
| Environmental conditions |
|
| Effect of temperature | Zero point: ≤ ± 0.0002 vol.% CO2/K; at 2.5 vol.% CO2: ≤ ± 0.0015 vol.% CO2/K |
| Effect of humidity | Zero point: ≤ 0.0001 vol.% CO2/% RH |
| Test gas | 2.0 vol.% CO2 in air. Before safety-critical measurements the calibration must be checked and adjusted if necessary. |
Response time
| Mode | Dräger X-am 5600 | Dräger X-am 8000 |
|---|---|---|
| Diffusion mode t50 | ≤ 15 seconds | ≤ 12 seconds |
| Diffusion mode t90 | ≤ 31 seconds | ≤ 50 seconds |
| Pump mode t50 | ≤ 8 seconds | ≤ 10 seconds |
| Pump mode t90 | ≤ 11 seconds | ≤ 15 seconds |
Main areas of application
- Carbon dioxide monitoring
- Industrial gas monitoring
- Applications requiring safe and stable CO2 measurement