How Dräger Tubes work

Dräger Tubes are a simple and reliable way to measure gases and vapours on site when you need a quick, selective result. They are especially useful for spot checks, clearance measurements and short-term investigations where a full electronic setup is not necessary.


How they work

  • Break off both ends of the glass tube.
  • Insert the tube into a compatible pump, with the arrow pointing towards the pump.
  • Draw a defined volume of air through the tube using the specified number of pump strokes.
  • The target gas reacts with the indicator layer inside the tube.
  • The colour change is read against the printed scale, or checked by colour comparison depending on the tube type.
  • The result shows the concentration at that exact place and time, so it gives you a clear snapshot of conditions on site.

This principle has stayed reliable for decades and follows the DIN EN ISO 17621 standard for short-term detector tube systems. Depending on the substance and tube type, measuring time typically ranges from around 10 seconds to 20 minutes, with an accuracy of roughly plus or minus 15 to 25 per cent. Because every result is a snapshot for that specific place and time, tubes are best suited to one-off answers rather than ongoing monitoring.


Where they are useful

Dräger Tubes are commonly used for workplace checks, maintenance tasks, clearance measurements before entry into vessels or confined spaces, and checks after an incident. For continuous monitoring with alarms, a Multigas Detector or Area Monitoring system is usually the better choice.


Choose the right setup for your application

Classic tube systems can be used with a manual pump or an automatic tube measuring device. For very low concentrations and highly selective measurements in the ppb range, Dräger MicroTubes provide the more advanced option and are evaluated with the X-act 7000. Several pre-layers inside each MicroTube filter out other substances before they reach the reactive layer, which significantly reduces cross-sensitivities and false-positive readings, even in complex mixtures.


A practical hire option

If you only need this type of measurement occasionally, hiring the measuring device and buying only the tubes you use is a cost-effective way to stay flexible for audits, one-off checks and short-term projects. Your hired equipment arrives maintained, checked and ready to use, so you can start measuring as soon as it is delivered.


Person mit gelbem Schutzhelm und blauer Arbeitskleidung bedient ein X-act 5000 Gerät, das mit einem Tank mit Schild Tank 2 verbunden ist
Person mit gelben Arbeitshandschuhen hält ein Messgerät und einen Verlängerungsschlauch von Dräger Accuro zur Gasmesstechnik


The expiry date is printed on every pack. Shelf life also depends on keeping the tubes within the storage temperature shown on the packaging. Once the expiry date has passed, the tube must no longer be used for a measurement.
No. Dräger Tubes are single-use consumables. The indicator layer inside the tube reacts chemically with the target gas, and this reaction cannot be reversed, so each tube can only deliver one measurement.
If you are not sure which gas or vapour you are dealing with, simultaneous test sets are the right choice. They combine five parallel Dräger Tubes that each indicate a different substance at once, giving you a quick first risk assessment before you decide on further steps.
Results can be recorded with the free Dräger-Tubes app for iOS and Android. It lets you scan the tube, log the reading together with photos and environmental data, and share the report by email, making documentation straightforward for audits.
Classic short-term tubes work with a manual pump or an automatic tube measuring device. Dräger MicroTubes are used exclusively with the X-act 7000. Some special tubes, such as PID pre-tubes, need a matching PID device rather than a standard tube pump.