The DrägerSensor XS EC H2O2 precisely measures the concentration of hydrogen peroxide and warns of hazardous gas concentrations.

  • Compatible with the Dräger X-am 5100
  • Expected service life of more than 2 years, supplied with a one-year warranty
  • Fast response time ≤ 60 seconds (t90), warm-up time ≤ 12 hours
  • Operating conditions: 0 to 50 °C, 10 to 90 % RH, 700 to 1,300 hPa

Technical data – DrägerSensor XS EC H2O2

Detection limit0.1 ppm
Resolution0.1 ppm
Measuring range0 to 20 ppm
Response time≤ 60 seconds (t90)
Accuracy (sensitivity)≤ ± 10 % of the measured value
Long-term drift at 20 °C (68 °F)
  • Zero point: ≤ ± 1 ppm/year
  • Sensitivity: ≤ ± 2 % of the measured value/month
Warm-up time≤ 12 hours
Environmental conditions
  • Temperature: 0 to 50 °C (32 to 122 °F)
  • Humidity: 10 to 90 % RH
  • Pressure: 700 to 1,300 hPa
Effect of temperature
  • Zero point: ≤ ± 1 ppm
  • Sensitivity: ≤ ± 0.5 % of the measured value/K
Effect of humidity
  • Zero point: ≤ ± 0.01 ppm/% RH
  • Sensitivity: ≤ ± 0.1 % of the measured value/% RH

Test gas: alternatively, the sensor can be calibrated with 10 ppm SO2. Such a surrogate calibration with SO2 may introduce an additional measurement error of up to 30 %. A function test must be carried out after a surrogate calibration or a sensor replacement; after 30 seconds at the latest, the reading must be greater than 1 ppm H2O2.

Relevant cross-sensitivities – DrägerSensor XS EC H2O2

The values given in the table are guide values and apply to new sensors. The values stated may vary by ±30%. The sensor may also be sensitive to other gases (data available from Dräger on request). Gas mixtures may be displayed as a combined total. Gases with a negative sensitivity can cancel out a positive H2O2 reading. It should therefore be checked whether gas mixtures are present.

Gas/vapourConcentrationReading in ppm H2O2
Acetylene (C2H2)200 ppm≤ 35
Chlorine (Cl2)5 ppm≤ 1 (−)
Hydrogen (H2)1.5 vol.%≤ 5
Hydrogen chloride (HCl)15 ppm≤ 3
Hydrogen cyanide (HCN)25 ppm≤ 7
Sulphur dioxide (SO2)20 ppm≤ 7

(−) negative sign of the deviation

Main areas of application

  • Disinfection and sterilisation
  • Bleaching processes
  • Decontamination of indoor spaces

Instructions for use

Further information can be found in the instructions for use: Instructions for use for this product