Pac 8000 Single Gas Monitor for Ammonia (NH3) US
The Pac 8000 NH3 gas warning device reliably and precisely detects the concentration of ammonia and warns of dangerous gas concentration.
- Single gas measurement device for personal measurement
- XXS sensor technology with high reaction speed and membrane filter
- Standard measurement range: 0 - 300 ppm, A1/A2 alarm thresholds: 20/40 ppm
- 360° alarm signal, acoustic (90 dB) and optical
- The gas detector features a robust, ergonomic housing for use under extreme conditions
- Large and user-friendly display with bright backlight
- Storage of concentrations and events with date and time
Conduct the required function tests before each use with Dräger test and calibration stations and the associated test gases. The basis for the function tests are the information sheets T021 / T023 (DGUV Information 213-056 or DGUV Information 213-057). Further information can be found here.
If you desire a custom alarm threshold configuration, please specify the alarm thresholds in the comment field at the end of the ordering process.
For more information about the product, visit our website. Please follow the link.
Technical Data - Dräger Sensor XXS NH3
| Detection Limit | 4 ppm |
|---|---|
| Resolution | 1 ppm |
| Measurement Range | 0 - 300 ppm NH3 (Ammonia) |
| Response Time | ≤ 10 seconds (t50) |
| Precision Sensitivity | ≤ ± 3% of measured value |
| Long-term Drift (at 20 °C) |
|
| Start-up Time | ≤ 12 hours |
| Environmental Conditions |
|
| Temperature Influence |
|
| Humidity Influence |
|
| Test Gas | approx. 10 to 75 ppm NH3 |
Relevant Cross Sensitivities - Dräger Sensor XXS NH3
The values indicated in the table are guideline values and apply to new sensors. The specified values may vary by ±30%. The sensor may also be sensitive to other gases (data available on request from Dräger). Gas mixtures may be displayed as a total. Gases with negative sensitivity can negate a positive NH3 display. It should be checked if gas mixtures are present.
| Gas/Vapor | Concentration | Display in ppm NH3 |
|---|---|---|
| Chlorine (Cl2) | 10 ppm | ≤ 30 (–) |
| Hydrochloric Acid (HCl) | 20 ppm | ≤ 15 (–) |
| Diethanolamine (C4H11NO2) | 10 ppm | 5 ppm |
| Ethanol (C2H5OH) | 250 ppm | ≤ 40 |
| Ethine (C2H2) | 100 ppm | no influence |
| Ethyldimethylamine (C4H11N) | 50 ppm | 30 ppm |
| Carbon Dioxide (CO2) | 10 vol.-% | no influence |
| Carbon Monoxide (CO) | 1000 ppm | no influence |
| Phosphine (PH3) | 1 ppm | ≤ 2 |
| Hydrogen Sulfide (H2S) | 20 ppm | ≤ 70 |
| Nitrogen Dioxide (NO2) | 20 ppm | ≤ 10 (–) |
| Nitric Oxide (NO) | 20 ppm | ≤ 10 |
| Hydrogen (H2) | 1000 ppm | ≤ 4 |
(-) negative sign of deviation